Kodak 8000c Service Manual [repack] -
External X-ray warning lamp terminals are standard. Emergency Stop: Features a red actuator push-button.
| Error Code | Meaning (From Service Manual) | Likely Fix | | :--- | :--- | :--- | | | Emergency stop activated (front panel or footswitch) | Check continuity of E-stop loop; replace microswitch. | | E-102 | KV (Kilovoltage) feedback out of range | High voltage cable arcing or faulty tube head. | | E-201 | mA (Milliamperage) filament circuit failure | Broken filament in X-ray tube or open circuit on filament driver board. | | E-301 | Rotation motor encoder failure | Clean encoder disc; replace optical sensor. | | E-512 | CCD communication timeout | Reseat flat flex cable to the sensor; check 5V rail. | kodak 8000c service manual